#メディカルテクニカ

email    contact

研究用 世界標準の
 #先天性心疾患用診断・経過・辞書の

#Pedcath8


#Pedcath8 のバージョン は、2015年新規ご購入分から適用されます。

現在、#Pedcath7 のご愛用の方は、#Pedcath8 へのバージョンアップで
ご変更させて頂きます。
その場合の費用は,平成27年6月1日メーカ訪問打ち合わせにより、
平成27年6月19日付けで、日本以外とは別途に、
低価格でメーカに設定頂きました。
お問い合わせ賜れば、ご案内申し上げます。

Pedcath8とPedcath7の主な違いは下記のようになっております。


#Labtech社 #wavelet_Algorithm #連続解析12誘導心電図ホルター
#P波検出能が有り、#心房細動などの
#自動解析検出能に優れる
#基線動揺、#ノイズ除去に優れたアルゴリズム採用

Theory of the #P_wave_detection

The algorithm first finds the  the possible positive and negative wave peaks based on zero transition searching, then validates them with comparing to reference P waves.

The P wave detection needs high amplitude resolution. This value is better, than 0.6 uV / bit in the Cardiospy system.  With this resolution and the effective filter system which uses wavelet transformation, the Cardiospy system is able to detect P waves less than 50 uV of amplitude.

 

Validation of the #P_wave_detector

The validation is carried out on 10 pcs 12 channel and 10 pcs 3 channel ECG reference records. The reference records include the P wave  annotation.  12 of the 20 records are taken from the MitBih database, 8 records are taken from the Labtech database (30000 – 30007). 

12 ch records

s0014lre, s0292lre, s0302lre, s0331lre, s0364lre, s0422_re, s0431_re, s0437_re, s0549_re, s0550_re

3 ch records

mgh001, mgh007, 30000, 30001, 30002, 30003, 30004, 30005, 30006, 30007

 

Validation result:

Sensitivity:                       95.42%

Positive predictivity:         97.16%


本邦においては現在 #研究向けに販売 2007年から下記モデル導入
#バイタルストリーム(#ケアテイカ_タイプ5の改名)
全面的グレードアップ #VitalStream_type1_type2_type3
#ケアテイカ_タイプ1_2_3_4、
#非観血血圧、#ワイヤレス、#連続、#ウエアラブル、#ブルーツース、
#観血血圧を、#非観血で、測定している事を検証され、#FDA認可取得(2017)、
全てバックアップをご用意、 国内は、薬事未承認


httpswww.ncbi.nlm.nih.govpmcarticlesPMC5361833

 

BMC #Anesthesiol. 2017; 17: 48.

Published online #2017 Mar 21. doi: 10.1186/s12871-017-0337-z

PMCID: PMC5361833

PMID: 28327093

#Continuous_Non-invasive_finger_cuff #CareTaker® comparable to #invasive_intra-arterial_pressure in patients undergoing major intra-abdominal surgery

Irwin Gratz,1 Edward Deal,1 Francis Spitz,1 Martin Baruch,2 I. Elaine Allen,3 Julia E. Seaman,4 Erin Pukenas,1 and Smith Jean1

Author information Article notes Copyright and License information Disclaimer

This article has been cited by other articles in PMC.

 

Associated Data

Data Availability Statement

The datasets generated during and analysed for the current study are available from the corresponding author on reasonable request.

 

Abstract

Background

Despite increased interest in non-invasive arterial pressure monitoring, the majority of commercially available technologies have failed to satisfy the limits established for the validation of automatic arterial pressure monitoring by the Association for the Advancement of Medical Instrumentation (AAMI). According to the ANSI/AAMI/ISO 81060–2:2013 standards, the group-average accuracy and precision are defined as acceptable if bias is not greater than 5 mmHg and standard deviation is not greater than 8 mmHg. In this study, these standards are used to evaluate the CareTaker® (CT) device, a device measuring continuous non-invasive blood pressure via a pulse contour algorithm called Pulse Decomposition Analysis.

Methods

A convenience sample of 24 patients scheduled for major abdominal surgery were consented to participate in this IRB approved pilot study. Each patient was monitored with a radial arterial catheter and CT using a finger cuff applied to the contralateral thumb. Hemodynamic variables were measured and analyzed from both devices for the first thirty minutes of the surgical procedure including the induction of anesthesia. The mean arterial pressure (MAP), systolic and diastolic blood pressures continuously collected from the arterial catheter and CT were compared. Pearson correlation coefficients were calculated between arterial catheter and CT blood pressure measurements, a Bland-Altman analysis, and polar and 4Q plots were created.

Results

The correlation of systolic, diastolic, and mean arterial pressures were 0.92, 0.86, 0.91, respectively (p<0.0001 for all the comparisons). The Bland-Altman comparison yielded a bias (as measured by overall mean difference) of −0.57, −2.52, 1.01 mmHg for systolic, diastolic, and mean arterial pressures, respectively with a standard deviation of 7.34, 6.47, 5.33 mmHg for systolic, diastolic, and mean arterial pressures, respectively (p<0.001 for all comparisons). The polar plot indicates little bias between the two methods (90%/95% CI at 31.5°/52°, respectively, overall bias=1.5°) with only a small percentage of points outside these lines. The 4Q plot indicates good concordance and no bias between the methods.

Conclusions

In this study, blood pressure measured using the non-invasive CT device was shown to correlate well with the arterial catheter measurements. Larger studies are needed to confirm these results in more varied settings. Most patients exhibited very good agreement between methods. Results were well within the limits established for the validation of automatic arterial pressure monitoring by the AAMI.

Keywords: Non-Invasive, CareTaker, Central blood pressure, Finger cuff, Intra-Arterial pressure

Go to:

Background

Accurate real-time continuous non-invasive blood pressure monitors (cNIBP) can bridge the gap between invasive arterial pressure monitoring and intermittent non-invasive sphygmomanometry. Latest developments in this field promise accuracy and the potential to lower risk and improve patient outcomes. However, a recent systematic review and meta-analysis of 28 studies using non-invasive technologies by Kim et al. reported that all failed to satisfy the limits that have been established for the validation of automatic arterial pressure monitoring by the Association for the Advancement of Medical Instrumentation (AAMI) [1]. According to this standard, the group-average accuracy and precision are defined as acceptable if bias is not greater than 5 mmHg and standard deviation is not greater than 8 mmHg. Kim et.al. obtained similar results when currently commercially available technologies were examined [1]. In addition, ease of use and patient comfort issues have been impediments to wider acceptance of current noninvasive cNIBP measurement methods. Their results suggest that currently available devices may not have the accuracy and precision for reliable clinical decisions, and there is a need for better devices.

We evaluated the CareTaker® (CT) device (Empirical Technologies Corporation, Charlottesville, Virginia) which has been described in detail elsewhere [2]. Briefly, the CT is a physiological sensing system that communicates physiological data wirelessly via Bluetooth (Fig. 1). The device uses a low pressure [35–45 mmHg], pump-inflated, cuff surrounding the proximal phalange of the thumb that pneumatically couples arterial pulsations via a pressure line to a custom-designed piezo-electric pressure sensor. This sensor converts the pressure pulsations, using transimpedance amplification, into a derivative voltage signal that is then digitized at 500 Hz, transmitted to and recorded on a computer.

 

The CT measures continuous noninvasive blood pressure via a pulse contour analysis algorithm called Pulse Decomposition Analysis (PDA) [3]. It is based on the concept that five individual component pressure pulses constitute the peripheral arterial pressure pulse. These component pulses are due to the left ventricular ejection and the reflections and re-reflections of the first component pulse from two central arteries reflection sites [2] [4]. The first reflection site is the juncture between thoracic and abdominal aorta, at the height of the renal arteries, while the second site arises from the interface between abdominal aorta and the common iliac arteries. The renal site reflects the pressure pulse because the juncture of the aortic arteries there features significant changes in arterial diameter and wall elasticity. The two reflected arterial component pressure pulses, the renal reflection pulse (P2) and the iliac reflection pulse (P3), counter-propagate with respect to the original pulse due to the left ventricular contraction (Fig. 2) and arrive in the arterial periphery, specifically at the radial or digital arteries, with distinct time delays [5]. The basic validity of the PDA model was recently corroborated in a detailed and comprehensive arterial tree numerical modeling analysis [6] that examined the effect of the different arterial segments of the central arteries, the iliac arteries and beyond on the pressure/flow pulse patterns in the digital arteries. The results clearly identified the central arterial reflection sites, as opposed to more distal sites, as being the primary contributors to the pulse patterns observed in the digits.

 

Quantification and validation of physiological parameters is accomplished by extracting pertinent component pulse parameters [7]. Since the device relies on pulse analysis to track blood pressure, the coupling pressure of the finger cuff is maintained constant and well below diastole, avoiding potential blood flow impediments.

The aim of the present study was to specifically compare the non-invasive arterial pressure values obtained with the CT to the reference invasive arterial pressure technique.

Go to:

Methods

The Cooper Health System Institutional Review Board approved the study, and all subjects gave informed written consent. Data from twenty-four adult patients requiring hemodynamic monitoring during major open abdominal surgery were analyzed in this study. Patients were not excluded due to other medical conditions.

Measurements were obtained during general anesthesia in these patients starting with induction. The induction of anesthesia was chosen because the blood pressure fluctuations and variability typically found during this period provided an opportunity to compare tracking accuracy under baseline and induced controlled dynamic conditions. The data was evaluated using the ANSI/AAMI/ISO 81060–2:2013-related standards of accuracy and precision [8].

Anesthesia procedure

After a stable signal was recorded, patients were induced under general anesthesia by using propofol (2-4 mg/kg) and fentanyl 250ug. Tracheal intubation was facilitated by the administration of rocuronium (0.6 mg/kg). Mechanical ventilation was started using a volume controlled ventilator to maintain an adequate saturation and an end-tidal carbon dioxide of 35 mmHg. Inhalational anesthetic (Isoflurane) was added to maintain a BIS monitoring of 40–45. Vasoactive drugs were used to maintain a MAP greater than 60 mmHg based on the catheter value. Hemodynamic variables were measured from both devices for the entire procedure. The MAP, systolic and diastolic blood pressures were continuously collected from the arterial catheter and CT and averaged over 10 s periods for both devices.

 

Invasive arterial pressure measurement

Standard arterial blood pressure monitoring was performed prior to the induction of anesthesia using a 20G intra-arterial catheter inserted in the radial artery under local anesthesia using ultra sound guidance. The catheter was connected to a disposable pressure transducer with standard low compliant tubing. The transducer was placed at heart level and zeroed to ambient pressure. The transducer data was digitized, processed and collected using the Datex-Ohmeda S/5 Collect system (Datex-Ohmeda Division, Instrumentarium Corporation, Helsinki, Finland). For analysis, MAP, systolic and diastolic blood pressures were averaged over 10 s intervals.

Non-invasive CareTaker arterial pulse signal recording

The arterial pressure pulse signal was continuously measured using the CT device. For this study the CT device was calibrated using the arterial line blood pressure, but calibration can also be based on non-invasive oscillometric or oscillometric/auscultatory measurements. A fifteen second window at the start of the 30 min overlap section was used to obtain an arterial stiffness reading averaged across 5 beats, which was then used to calculate the PDA parameters for the blood pressure conversions (Fig. 2). With the exception of the four cases mentioned above, patient-specific PDA parameters, once established, were not changed for the matching procedure, irrespective of arterial stiffness or heart rate changes. On four occasions for the entire data set, the offsets of the linear conversion equations were changed as a result of persistent changes in arterial stiffness or heart rate changes exceeding 30%. The PDA algorithm has recently been validated and described elsewhere [6].



動脈硬化指標、ABPM(収縮期、拡張期、

と24時間から72時間血圧値を測定します。薬事認可有り、日本特許有り


アルテリオグラフフ24 中心血圧及び関連パラメータは検証対象です

 
Medical Teknika MedicalTeknika Medical Teknika Medical Teknika Medical Teknika
Medical Teknika Medical Teknika Medical Teknika Medical Teknika Medical Teknika
Medical Teknika Medical Teknika Medical Teknika Medical Teknika Medical Teknika
Medical Teknika Medical Teknika Medical Teknika Medical Teknika Medical Teknika
お問い合わせ先 メディカルテクニカ有限会社
#南魚沼市民病院ーCDC #メディカルテクニカ掲示板 #メディカルテクニカ掲示板 #輸入代行
#Angioscopy #Radiofrequency #先生のホームページ掲載
リンク リンク 関係先 ナノテク材料
#メンネンメディカル #ウエアラブル12誘導心電図電極 #西陣織12誘導心電図電極 #ワンタッチ心電図電極
#ウエアラブルモバイル心電図 #眼鏡にPC画面表示 SHARPロボット #ShimmerSensingにLabview
ウエアラブル式胸部12誘導心電図 多種類センサー式ワイヤレスアンプ #ワイヤレス用USB ウエアラブルの米国例
ウエアラブル米国例 第三回2017ウエアラブル展より ウエアラブルセンサー例 第二回2016ウエアラブル展より
スマートめがね ウエアラブル用素材 スマホ装着生体プリアンプ
エアロビック モバイル精神分析 自律神経解析 神経伝導速度検査
リハビリ画像解析 神経信号伝達 高度自律神経解析 電気刺激
糖尿病診断 脊髄機能補完 コスモス製リハ用具 超音波血流計
水中浮遊着衣センサ付き 国産ホルタ電極 自律神経解析ソフト 自律神経解析ソフト
エネルギー代謝 めまい治療 筋電刺激 脳刺激マッピング
耳栓検知生体信号モニタ
お問い合わせ先 #メディカルテクニカ有限会社
新型血圧測定 #ケアテイカメディカル
#VitalStream_type1
#ケアテイカ
#VitaStream_type2
血圧制御
#VitalStream_type3
#ケアテイカ
#VitalStream_caretaker
#ケアテイカ案内
#VitalStream_Cardiac_Surgery
#ケアテイカのカフの位置
#VitalStream_Hypertension
体内血圧測定
#MRI下非観血連続血圧計
アルテリオグラフの中心血圧測定 オーグメンテイション フレイルティメータ 中心血検証をお願いします
アルテリオグラフの文献例
聴診等用標準器 医療機器校正器類 カルディオニクス製製品 カルディオニクス製シミュレータ
画像表示付き聴診器 医療用標準器 ワイヤレス校正器 同時聴診教育システム
モバイル聴診 聴診音同時多人数 電子聴診器のソフト 各種電子聴診器
モバイル聴診 血管狭窄診断聴診器 聴診音画像化聴診器 タイマ無し出力付き研究電子聴診器
遠隔血栓検知
#先天性心疾患ソフト Pedcath7 #Pedcath8 川崎病
Pedcath概要 Pedcathの本 Pedcathの仲間 Pedcath選定理由書
Pedcath参考画像 Pedcath品目 Pedcathマルティユーザ Pedcath拡張機能
佐野シャント ペドカスコンパニオン タブレット補助Pedcath
ラブテック社心臓リハビリ メタボリックテスト ラブテック社ワイヤレス12誘導心電計 ラブテック社12誘導心電計
ラブテック12誘導心電計 心電図解析 ラブテック社モバイル心電計 検診用心電計
#ラブテックホルタ心電計
#VectorECG
#ラブテックホルタ心電計
#VectorECG
#ラブテックホルタ心電計
#ベクトル心電図
#ラブテックホルタ心電計
#心房細動自動検出
#ラブテックホルタ心電計
#心房細動自動検出
#ラブテックネット心電図 シムレータ
ラブテック社心臓リハビリ
#ラブテックホルタ心電計資料
#PwaveAutoDetect
心臓リハビリ12誘導心電計 タブレットの12誘導心電図での利用例
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link