|
Theory of the #P_wave_detectionThe algorithm first
finds the the possible positive and negative wave peaks
based on zero transition searching, then validates them with comparing
to reference P waves. The P wave detection needs high amplitude resolution. This value is better,
than 0.6 uV / bit in the Cardiospy system. With this resolution and
the effective filter system which uses wavelet transformation,
the Cardiospy system is able to detect P waves less than 50 uV of amplitude. Validation of the P wave detectorThe validation is carried out on 10 pcs 12 channel and 10 pcs 3 channel
ECG
reference records. The reference records include the P wave
annotation. 12
of the 20 records are taken from the MitBih database, 8 records are
taken from the Labtech database (30000 ? 30007).
12 ch recordss0014lre, s0292lre, s0302lre, s0331lre, s0364lre, s0422_re, s0431_re,
s0437_re, s0549_re, s0550_re3 ch recordsmgh001, mgh007,
30000, 30001, 30002, 30003, 30004, 30005, 30006, 30007 Validation result:Sensitivity: 95.42%
Positive
predictivity: 97.16%
|