#メディカルテクニカ

email     contact

#P波自動検出能で、#心房細動自動解析を
ご評価頂き学術文献ご発表なさる先生
をお待ち申し上げております


#ラブテック ホルターの解析機能として

#Heart_Rate_Variability ,#Turbulence、#3D可変表示、
#Vector_Cardiolography, #Spectral_Analysis 等が装備されています

#認知症、#痴呆症、#アルツファイマー、#脳卒中、#COPD、#腎疾患、#睡眠障害
#高脂血症、#高血圧、#肥満、#生活習慣病、#腎臓病、#内分泌など




#Atrial_fibrillation

Atrial fibrillation (AF) is the most common arrhythmia lasting for more than 30 seconds. Its prevalence in the population increases with age, and it is estimated to affect over 4 percent of the population above the age of 60. Patients with atrial fibrillation are at an increased risk of ischemic stroke and other thromboembolic events.

The electrocardiogram (ECG) is used to verify the presence of AF and is necessary to make the diagnosis. Holter monitoring or event recorders are used to identify the arrhythmia if it is intermittent and not captured on routine electrocardiography and assess overall ventricular response rates. Longer monitoring terms using external or implantable recorders, multiple daily ECG checks for a prolonged period and continuous telemetry are the leading diagnostic ECG methods to identify subclinical cases of intermittent (paroxysmal) AF.

The electrocardiogram in a patient with AF has the following main characteristics:

1.       Absence of discrete P waves

2.       Fibrillatory or f waves are present at a rate that is generally between 350 and 600 beats/minute; the f waves vary in amplitude, morphology, and intervals

3.       The RR intervals follow no repetitive pattern; they have been labelled as “irregularly irregular.”

4.       The QRS complexes are narrow unless AV conduction is abnormal because of other condition

Labtech Cardiospy® uses the first and third characteristics to identify sections of atrial fibrillation on Holter ECG records. An ECG section is only marked as atrial fibrillation if the RR intervals are irregularly irregular and there are no discrete P-waves at predefined locations before QRS complexes. Fibrillatory f waves are often hard to ascertain on Holter ECG records, therefore they are not used as a requirement for the detection of AF. Manual corrections can easily be done after the automatic analysis if necessary, but our internal tests show that the accuracy of the automatic AF detection exceeds 90%.


   The red strip marks the automatically detected AF sections

(2)    The NN intervals plot helps the user to assess the irregularity                           a) AF section          b) normal section

(3)    ECG strip with markers for atrial fibrillation


様々な #電子聴診器や #単品機能品など
 #代行輸入申し上げます

 診療報酬有 #動脈硬化指標を24時間から72時間トレンド計測
 
#三分間で測定 検証対象の #中心血圧測定
#欧州米国特許有 #日本特許有 携帯電話及びインターネット利用、
#認知症、#痴呆症、#アルツファイマー、#脳卒中、#COPD、#腎疾患、#睡眠障害
#高脂血症、#高血圧、#肥満、#生活習慣病、#腎臓病、#内分泌など



#米国製 #Caretaker_Medical 社は、
#世界初 #非観血式計測で、#連続観血血圧解析し、
#生体情報の #血行動態 #パラメータを20種類連続解析
#世界最先端ディジタル搭載し、#追加オプション可能な設計
#2017年FDA認可、2024年世界麻酔学会が検証しています
#メディカルテクニカは2007年から研究向けに導入

httpswww.ncbi.nlm.nih.govpmcarticlesPMC5361833

 

BMC Anesthesiol. 2017; 17: 48.

Published online 2017 Mar 21. doi: 10.1186/s12871-017-0337-z

PMCID: PMC5361833

PMID: 28327093

Continuous Non-invasive finger cuff CareTaker® comparable to invasive intra-arterial pressure in patients undergoing major intra-abdominal surgery

Irwin Gratz,1 Edward Deal,1 Francis Spitz,1 Martin Baruch,2 I. Elaine Allen,3 Julia E. Seaman,4 Erin Pukenas,1 and Smith Jean1

Author information Article notes Copyright and License information Disclaimer

This article has been cited by other articles in PMC.

 

Associated Data

Data Availability Statement

The datasets generated during and analysed for the current study are available from the corresponding author on reasonable request.

 

Abstract

Background

Despite increased interest in non-invasive arterial pressure monitoring, the majority of commercially available technologies have failed to satisfy the limits established for the validation of automatic arterial pressure monitoring by the Association for the Advancement of Medical Instrumentation (AAMI). According to the ANSI/AAMI/ISO 81060–2:2013 standards, the group-average accuracy and precision are defined as acceptable if bias is not greater than 5 mmHg and standard deviation is not greater than 8 mmHg. In this study, these standards are used to evaluate the CareTaker® (CT) device, a device measuring continuous non-invasive blood pressure via a pulse contour algorithm called Pulse Decomposition Analysis.

Methods

A convenience sample of 24 patients scheduled for major abdominal surgery were consented to participate in this IRB approved pilot study. Each patient was monitored with a radial arterial catheter and CT using a finger cuff applied to the contralateral thumb. Hemodynamic variables were measured and analyzed from both devices for the first thirty minutes of the surgical procedure including the induction of anesthesia. The mean arterial pressure (MAP), systolic and diastolic blood pressures continuously collected from the arterial catheter and CT were compared. Pearson correlation coefficients were calculated between arterial catheter and CT blood pressure measurements, a Bland-Altman analysis, and polar and 4Q plots were created.

Results

The correlation of systolic, diastolic, and mean arterial pressures were 0.92, 0.86, 0.91, respectively (p<0.0001 for all the comparisons). The Bland-Altman comparison yielded a bias (as measured by overall mean difference) of −0.57, −2.52, 1.01 mmHg for systolic, diastolic, and mean arterial pressures, respectively with a standard deviation of 7.34, 6.47, 5.33 mmHg for systolic, diastolic, and mean arterial pressures, respectively (p<0.001 for all comparisons). The polar plot indicates little bias between the two methods (90%/95% CI at 31.5°/52°, respectively, overall bias=1.5°) with only a small percentage of points outside these lines. The 4Q plot indicates good concordance and no bias between the methods.

Medical Teknika MedicalTeknika Medical Teknika Medical Teknika Medical Teknika
Medical Teknika Medical Teknika Medical Teknika Medical Teknika Medical Teknika
Medical Teknika Medical Teknika Medical Teknika Medical Teknika Medical Teknika
Medical Teknika Medical Teknika Medical Teknika Medical Teknika Medical Teknika
お問い合わせ先 メディカルテクニカ有限会社
新型血圧測定 #ケアテイカメディカル
#VitalStream_type1
#ケアテイカ
#VitaStream_type2
血圧制御
#VitalStream_type3
#ケアテイカ
#VitalStream_caretaker
#ケアテイカ案内
#VitalStream_Cardiac_Surgery
#ケアテイカのカフの位置
#VitalStream_Hypertension
体内血圧測定
#MRI下非観血連続血圧計
アルテリオグラフの中心血圧測定 オーグメンテイション フレイルティメータ 中心血検証をお願いします
アルテリオグラフの文献例
聴診等用標準器 医療機器校正器類 カルディオニクス製製品 カルディオニクス製シミュレータ
画像表示付き聴診器 医療用標準器 ワイヤレス校正器 同時聴診教育システム
モバイル聴診 聴診音同時多人数 電子聴診器のソフト 各種電子聴診器
モバイル聴診 血管狭窄診断聴診器 聴診音画像化聴診器 タイマ無し出力付き研究電子聴診器
遠隔血栓検知
#先天性心疾患ソフト Pedcath7 #Pedcath8 川崎病
Pedcath概要 Pedcathの本 Pedcathの仲間 Pedcath選定理由書
Pedcath参考画像 Pedcath品目 Pedcathマルティユーザ Pedcath拡張機能
佐野シャント ペドカスコンパニオン タブレット補助Pedcath
ラブテック社心臓リハビリ メタボリックテスト ラブテック社ワイヤレス12誘導心電計 ラブテック社12誘導心電計
ラブテック12誘導心電計 心電図解析 ラブテック社モバイル心電計 検診用心電計
#ラブテックホルタ心電計
#VectorECG
#ラブテックホルタ心電計
#VectorECG
#ラブテックホルタ心電計
#ベクトル心電図
#ラブテックホルタ心電計
#心房細動自動検出
#ラブテックホルタ心電計
#心房細動自動検出
#ラブテックネット心電図 シムレータ
ラブテック社心臓リハビリ
#ラブテックホルタ心電計資料
#PwaveAutoDetect
心臓リハビリ12誘導心電計 タブレットの12誘導心電図での利用例
ワイヤレスブルーツース100mの例
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link