|   | Theory of the #P_wave_detectionThe algorithm first
finds the  the possible positive and negative wave peaks based on zero transition searching, then validates them with comparing
 to reference P waves.
 The P wave detection needs high amplitude resolution. This value is better,
            than 0.6 uV / bit in the Cardiospy system.  With this resolution and
 the effective filter system which uses wavelet transformation,
 the Cardiospy system is able to detect P waves less than 50 uV of amplitude.
   Validation of the P wave detector The validation is carried out on 10 pcs 12 channel and 10 pcs 3 channel
            ECGreference records. The reference records include the P wave 
annotation.  12
 of the 20 records are taken from the MitBih database, 8 records are
 taken from the Labtech database (30000 ? 30007).
 12 ch records s0014lre, s0292lre, s0302lre, s0331lre, s0364lre, s0422_re, s0431_re,s0437_re, s0549_re, s0550_re
 3 ch records mgh001, mgh007,
30000, 30001, 30002, 30003, 30004, 30005, 30006, 30007   Validation result: Sensitivity:                       95.42% Positive
predictivity:         97.16% |