email     contact

エアロビック運動が、
慢性心不全疾患患者にとって、
最善のフィットネス
リハビリ方法と証明




PCの機能が向上し、#Descrete_Wavelet 手法が、お求めやすい低価格で
実現し、#心房のP波を自動検出できるようになりました、
#心房細動の自動解析能をお試し賜われれば幸いです
臨床用 薬事認可済


#Labtech社は、当社独自の方法による
#P波自動検出と #心房細動自動解析手法
#T_Wave_Alternans 解析手法を開発し、当社 #ホルターに搭載。


#VectorCardiology #Heart_Rate_Variability , #TWA_オルタナンス
#Turbulence、#3D可変表示、#Spectral_Analysis等が装備

Theory of the #P_wave_detectionThe algorithm first finds the  the possible positive and negative wave peaks
based on zero transition searching, then validates them with comparing
to reference P waves.
The #P_wave_detection needs high amplitude resolution. This value is better,
than 0.6 uV / bit in the Cardiospy system.  With this resolution and
the effective filter system which uses wavelet transformation,
the Cardiospy system is able to detect P waves less than 50 uV of amplitude.
 Validation of the #P_wave_detectorThe validation is carried out on 10 pcs 12 channel and 10 pcs 3 channel ECG
reference records. The reference records include the P wave  annotation.  12
of the 20 records are taken from the MitBih database, 8 records are
taken from the Labtech database (30000 ? 30007). 
12 ch records s0014lre, s0292lre, s0302lre, s0331lre, s0364lre, s0422_re, s0431_re,
s0437_re, s0549_re, s0550_re
3 ch recordsmgh001, mgh007, 30000, 30001, 30002, 30003, 30004, 30005, 30006, 30007 Validation result:Sensitivity:                       95.42% Positive predictivity:         97.16%

#ケアテイカメディカル #caretaker_Medical社 #2017年FDA認可ー #本邦研究用
#2007年から、#メディカルテクニカは #バックアップをご用意して導入
#Caretaker_type1_type2_type3_type4
#VitalStream_type1_type2_type3
#非観血血圧計測で・#連続観血血圧解析・ #顧客の仕様追加に対応できる設計
・ #ポータブル・ #ウエアラブル・ #一拍毎解析出力付き

httpswww.ncbi.nlm.nih.govpmcarticlesPMC5361833

 

#BMC_Anesthesiol. 2017; 17: 48.

Published online 2017 Mar 21. doi: 10.1186/s12871-017-0337-z

PMCID: PMC5361833

PMID: 28327093

#Continuous_Non-invasive_finger_cuff_CareTakerR comparable to #invasive_intra-arterial_pressure in patients undergoing major intra-abdominal surgery

Irwin Gratz,1 Edward Deal,1 Francis Spitz,1 Martin Baruch,2 I. Elaine Allen,3 Julia E. Seaman,4 Erin Pukenas,1 and Smith Jean1

Author information Article notes Copyright and License information Disclaimer

This article has been cited by other articles in PMC.

 

Associated Data

Data Availability Statement

The datasets generated during and analysed for the current study are available from the corresponding author on reasonable request.

 

Abstract

Background

Despite increased interest in non-invasive arterial pressure monitoring, the majority of commercially available technologies have failed to satisfy the limits established for the validation of automatic arterial pressure monitoring by the Association for the Advancement of Medical Instrumentation (AAMI). According to the ANSI/AAMI/ISO 81060?2:2013 standards, the group-average accuracy and precision are defined as acceptable if bias is not greater than 5 mmHg and standard deviation is not greater than 8 mmHg. In this study, these standards are used to evaluate the CareTakerR (CT) device, a device measuring continuous non-invasive blood pressure via a pulse contour algorithm called Pulse Decomposition Analysis.

Methods

A convenience sample of 24 patients scheduled for major abdominal surgery were consented to participate in this IRB approved pilot study. Each patient was monitored with a radial arterial catheter and CT using a finger cuff applied to the contralateral thumb. Hemodynamic variables were measured and analyzed from both devices for the first thirty minutes of the surgical procedure including the induction of anesthesia. The mean arterial pressure (MAP), systolic and diastolic blood pressures continuously collected from the arterial catheter and CT were compared. Pearson correlation coefficients were calculated between arterial catheter and CT blood pressure measurements, a Bland-Altman analysis, and polar and 4Q plots were created.

Results

The correlation of systolic, diastolic, and mean arterial pressures were 0.92, 0.86, 0.91, respectively (p?<?0.0001 for all the comparisons). The Bland-Altman comparison yielded a bias (as measured by overall mean difference) of ?0.57, ?2.52, 1.01 mmHg for systolic, diastolic, and mean arterial pressures, respectively with a standard deviation of 7.34, 6.47, 5.33 mmHg for systolic, diastolic, and mean arterial pressures, respectively (p?<?0.001 for all comparisons). The polar plot indicates little bias between the two methods (90%/95% CI at 31.5°/52°, respectively, overall bias?=?1.5°) with only a small percentage of points outside these lines. The 4Q plot indicates good concordance and no bias between the methods.

Conclusions

In this study, #blood_pressure measured using the #non-invasive_CT_device was shown to correlate well with the #arterial_catheter_measurements. Larger studies are needed to confirm these results in more varied settings. Most patients exhibited very good agreement between methods. Results were well within the limits established for the validation of #automatic_arterial_pressure_monitoring by the AAMI.

Keywords: Non-Invasive, CareTaker, Central blood pressure, Finger cuff, Intra-Arterial pressure


#Tensiomed 社 #Arteriograph 24 #血管脈波検査装置
ご注意:中心血圧と関連機能は検証が必要です、

#臨床用ー #薬事認可済

Medical Teknika MedicalTeknika Medical Teknika Medical Teknika Medical Teknika
Medical Teknika Medical Teknika Medical Teknika Medical Teknika Medical Teknika
Medical Teknika Medical Teknika Medical Teknika Medical Teknika Medical Teknika
Medical Teknika Medical Teknika Medical Teknika Medical Teknika Medical Teknika
お問い合わせ先 メディカルテクニカ有限会社
#南魚沼市民病院ーCDC #メディカルテクニカ掲示板 #メディカルテクニカ掲示板 #輸入代行
#Angioscopy #Radiofrequency #先生のホームページ掲載
リンク リンク 関係先 ナノテク材料
#メンネンメディカル #ウエアラブル12誘導心電図電極 #西陣織12誘導心電図電極 #ワンタッチ心電図電極
#ウエアラブルモバイル心電図 #眼鏡にPC画面表示 SHARPロボット #ShimmerSensingにLabview
ウエアラブル式胸部12誘導心電図 多種類センサー式ワイヤレスアンプ #ワイヤレス用USB ウエアラブルの米国例
ウエアラブル米国例 第三回2017ウエアラブル展より ウエアラブルセンサー例 第二回2016ウエアラブル展より
スマートめがね ウエアラブル用素材 スマホ装着生体プリアンプ
エアロビック モバイル精神分析 自律神経解析 神経伝導速度検査
リハビリ画像解析 神経信号伝達 高度自律神経解析 電気刺激
糖尿病診断 脊髄機能補完 コスモス製リハ用具 超音波血流計
水中浮遊着衣センサ付き 国産ホルタ電極 自律神経解析ソフト 自律神経解析ソフト
エネルギー代謝 めまい治療 筋電刺激 脳刺激マッピング
耳栓検知生体信号モニタ
お問い合わせ先 #メディカルテクニカ有限会社
新型血圧測定 #ケアテイカメディカル
#VitalStream_type1
#ケアテイカ
#VitaStream_type2
血圧制御
#VitalStream_type3
#ケアテイカ
#VitalStream_caretaker
#ケアテイカ案内
#VitalStream_Cardiac_Surgery
#ケアテイカのカフの位置
#VitalStream_Hypertension
体内血圧測定
#MRI下非観血連続血圧計
アルテリオグラフの中心血圧測定 オーグメンテイション フレイルティメータ 中心血検証をお願いします
アルテリオグラフの文献例
聴診等用標準器 医療機器校正器類 カルディオニクス製製品 カルディオニクス製シミュレータ
画像表示付き聴診器 医療用標準器 ワイヤレス校正器 同時聴診教育システム
モバイル聴診 聴診音同時多人数 電子聴診器のソフト 各種電子聴診器
モバイル聴診 血管狭窄診断聴診器 聴診音画像化聴診器 タイマ無し出力付き研究電子聴診器
遠隔血栓検知
#先天性心疾患ソフト Pedcath7 #Pedcath8 川崎病
Pedcath概要 Pedcathの本 Pedcathの仲間 Pedcath選定理由書
Pedcath参考画像 Pedcath品目 Pedcathマルティユーザ Pedcath拡張機能
佐野シャント ペドカスコンパニオン タブレット補助Pedcath
ラブテック社心臓リハビリ メタボリックテスト ラブテック社ワイヤレス12誘導心電計 ラブテック社12誘導心電計
ラブテック12誘導心電計 心電図解析 ラブテック社モバイル心電計 検診用心電計
#ラブテックホルタ心電計
#VectorECG
#ラブテックホルタ心電計
#VectorECG
#ラブテックホルタ心電計
#ベクトル心電図
#ラブテックホルタ心電計
#心房細動自動検出
#ラブテックホルタ心電計
#心房細動自動検出
#ラブテックネット心電図 シムレータ
ラブテック社心臓リハビリ
#ラブテックホルタ心電計資料
#PwaveAutoDetect
心臓リハビリ12誘導心電計 タブレットの12誘導心電図での利用例
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link
Medical Teknika add link Medical Teknika add link Medical Teknika add link Medical Teknika add link